
Clinical Window Educational Program
is sponsored by GE Healthcare

Clinical Window Web Journal
complies with the
HON code standard for
trustworthy health information:
verify
here.
Educational collaboration:

Clinical Window
is main sponsor for this concise, textbook style website of anaesthesiology
|
 |
Nocturnal Pulse Oximetry Studies in Patients
with COPD
Part II: Treatment of nocturnal hypoxemia and
practical case examples
 |
Ari Chaouat, MD
Service de Pneumologie
Hôpital de Hautepierre
Strasbourg Cedex
France
|
The article also available in PDF:
195KB
Treatment of nocturnal hypoxemia in COPD
One of the prime elements in the treatment of advanced
chronic obstructive pulmonary disease (COPD) accompanied with respiratory
failure is to correct hypoxemia. While the indications for standard oxygen
therapy are well defined [28], the same cannot be said of isolated nocturnal
oxygen therapy. There is still controversy as to the importance of treating
such COPD patients who are not severely hypoxemic in daytime, but who
experience marked nighttime desaturation during sleep.
Standard long-term oxygen therapy
is indicated for COPD patients with marked, persistent hypoxemia, a daytime
PaO2 below 55-60 mmHg. This is the important group of patients who demonstrate
the most profound nocturnal hypoxemia. To be effective, long-term oxygen
therapy should be applied for at least 16 hours out of 24 hours, and the
sleep period must be totally covered [28].
The usual rate of oxygen administration is 1.5 to
3 L/min. One may ask are these rates sufficient to correct episodes of
potentially severe desaturation during sleep? In a landmark study, an
oxygen flow rate of 2 L/min was seen to be effective, as the mean SpO2
was improved from 53 ±29 % to 90 ±9% [29]. Episodes of desaturation
still persisted, but there were less of them and they were not that severe.
The efficacy of standard oxygen therapy at a normal
rate has been confirmed by other studies, as well [6,21,30], However,
nocturnal oximetry is useful in checking that the selected oxygen flow
will ensure nighttime SpO2 of over 90%. In general, nocturnal oxygen therapy
has a favorable effect on pulmonary hypertensive jolts [21] and on cardiac
rhythm disorders [23].
In oxygen therapy, one might fear of a progressive
rise in PaCO2 during sleep, as hypoxic stimulus is suppressed, and there
may be a diminished ventilatory response to CO2. Several studies have
demonstrated that the risk is minimal, at least in the case of stable
patients with COPD [29,30]. However, obstructive sleep apnea syndrome
(OSAS) is worth noting here, as well a subgroup of patients with COPD
but without OSAS. They may develop a progressive increase in PaCO2, and
often experience acute exacerbation of the disease under oxygen therapy.
They may benefit from long-term ventilatory support.
It is good to bear in mind that patients with COPD
and OSAS may show a greater rise in PaCO2, which emphasizes importance
of making correct diagnosis at an early stage.
Oxygen therapy only during sleep:
Some COPD patients whose condition does not indicate standard, long-term
oxygen therapy may still suffer severe nighttime hypoxemia. A study with
40 COPD patients having daytime PaO2 of 60 to 65 mmHg showed that 18 of
40 spent over 30% of their sleep time with a SpO2 below 90% [8].
Nocturnal oxygen therapy might be indicated, but there
is not enough proof regarding the effect of sleep time oxygen therapy
for patients with COPD [31]. More studies are needed on the effects of
isolated nocturnal hypoxemia, as well.
Practicalities of nocturnal SpO2 monitoring
Pulse oximeters are invaluable, non-invasive tools
for the assessment of nocturnal hypoxemia in patients with COPD. The two
most common events that interfere with accurate measurement of SpO2, motion
and low perfusion, are generally not encountered in sleep studies for
patients with COPD. Indeed, SpO2 monitoring of is performed during a stable
state of the disease, and patients who are agitated or in a state of shock
are excluded.
Provided that the patient is cooperative, continuous
nocturnal SpO2monitoring is usually accurate. However, due to the high
proportion of smokers in the COPD population, it could be beneficial to
check the amount of carbon monoxide with a whole-blood oximeter just before
starting the sleep study. Benefits and cautions regarding nocturnal pulse
oximetry in patients with COPD are summarized in Table 1.
Table 1: Practical aspects related
to nocturnal pulse oximetry in patients with COPD.
| Benefits of pulse oximetry |
Points to note |
| Non invasive and well tolerated |
Check carbon monoxide (CO) levels to ensure SpO2
accuracy with smokers |
| Monitoring practically in real-time |
Accuracy of devices at low SpO2 values may be
variable |
| Can assess the effect of oxygen therapy |
Clip sensors may be dislodged during sleep, adhesive
sensors may provide more secure connections |
| Results may suggest patients requiring polysomnography |
|
Clinical cases
Case 1, diagnosis of sleep-related hypoxemia
in COPD.
Traditionally, sleep studies in COPD patients have mostly been recommended
only in two scenarios:
- Hypoxemia or right heart failure develops in the presence of relatively
mild airflow limitation
- The patient has symptoms suggesting the presence of sleep apnea
Despite that rather narrow definition, this author
feels that nighttime monitoring of SpO2 is indicated in all patients with
a resting daytime PaO2 of below 65 mmHg.
Figure 2 included in this part II of the article,
shows examples of continuous measurements of SpO2 in two patients with
COPD. Both patients had daytime PaO2 of 64 mmHg. Please note the lower
trend tracing, as it is representative of a typical nocturnal "desaturator"
patient. The patient spent 155 minutes of the recording with a SpO2 below
90 %.

Figure 2. Two patient examples
(A and B) with different over night pulse oximetry trends. Both patients
had daytime PaO2 of 64 mmHg: Trend tracing in A is characteristic to a
"non-desaturator", whereas curve B is a typical example of a
"desaturator".
Case 2, Diagnosis of associated OSAS.
When obstructive sleep apnea syndrome is suspected, a continuous nighttime
monitoring of SpO2 should be performed. Such a recording may show profound
hypoxemia (Figure 3), particularly during REM sleep. In addition, there
may be a characteristic swinging SpO2 pattern seen outside REM sleep.
This phenomenon is often related to respiratory obstructive events. The
suspected diagnosis of OSAS must then be confirmed with a polysomnography.

Figure 3. Over night oximetry
of a patient having an association of COPD and OSAS. The above (red) tracing
shows profound hypoxemia, particularly during REM sleep. There are the
characteristic swinging saturation patterns even outside REM sleep. This
is related to obstructive apnea and obstructive hypopnea.
Case 3: Control of the level of SpO2 under long-term oxygen therapy or
long-term ventilatory support.
The primary goal of oxygen therapy and ventilatory support is to increase
the baseline PaO2 to over 60 mmHg (8.0 kPa), or to produce a SpO2 of over
90 %: Hence, monitoring SpO2 during sleep would be advisable.
A 62 year-old male with severe COPD was a candidate
for long-term oxygen therapy. This patient’s typical pulmonary physiologic
values are shown in table 2.
Table 2: Characteristics of the patient
with severe COPD.
- Male, 62 year-old, ex-smoker (50 pack-year)
- Severe dyspnea on exertion
- FEV1 (23 % of the predicted value) 0.67 L
- FVC (38 % of the predicted value) 1.41 L
- FEV1/FVC -ratio 47 %
- Arterial oxygen tension (PaO2) 4 8 mm Hg
- Arterial carbon dioxide tension (PaCO2) 58 mm Hg
- FEV1 = Forced expiratory volume in 1 sec.
- FVC = Forced vital capacity
Initially, therapy was initiated with oxygen flow
of 3 L/min, allowing a satisfactory improvement of hypoxemia at rest,
but only in daytime. Despite, the nocturnal oximetry recording showed
constant low mean nighttime SpO2 values of 83 %.
Due to frequent exacerbations of COPD, an elevated
PaCO2 and persistent sleep time desaturation during oxygen therapy, long-term
ventilatory support was started in addition to the oxygen therapy. Thereafter,
nighttime recordings of SpO2 showed improvement to a mean SpO2 of 92 %.
Notably, the patient spent only 2 % of the monitored time with SpO2 values
below 90%.
Figure 4. Oximetry during
sleep in a patient with very severe COPD, graph illustrate three different
treatment periods. Periods A shows SpO2 with ambient air, whereas B was
during oxygen therapy (3 L/min), and C when patient received both ventilatory
support and oxygen therapy. As the current recording shows, SpO2 was satisfactory
(> 90 %) only during period C.
The recordings in figure 4 were performed during the
same night, 24 hours before patient’s discharge. Three weeks had
elapsed after admission for acute exacerbation of COPD.
There were three, 30 minute sleep periods: under ambient
air, under oxygen therapy, and under ventilatory support plus oxygen therapy.
The recordings demonstrate a gradual increase in SpO2 while new modalities
of therapy were initiated. Such a SpO2 recording is very practical in
controlling the therapeutic oxygen flow rate, as well as for adjusting
the patient’s ventilator pressure and volume settings.
This case demonstrates that nocturnal SpO2 monitoring
was effective in facilitating the choice of care in chronic respiratory
failure, and allowed a suitable correction of nocturnal hypoxemia (Figure
4).
References (parts I and II combined)
- Trask CH, Cree EM. Oximeter studies on patients with chronic obstructive
emphysema, awake and during sleep. N Engl J Med 1962;266:639-642.
- Koo KW, Sax DS, Snider GL. Arterial blood gases and pH during sleep
in chronic obstructive pulmonary disease. Am J Med 1975;58:663-670.
- Flick MR, Block AJ. Continuous in-vivo monitoring of arterial oxygenation
in chronic obstructive lung disease. Ann Intern Med 1977;86:725-730.
- Wynne JW, Block AJ, Hemenway J, Hunt LA, Flick MR. Disordered breathing
and oxygen desaturation during sleep in patients with chronic obstructive
lung disease (COLD). Am J Med 1979;66:573-579.
- Catterall JR, Douglas NJ, Calverley PM, Shapiro CM, Brezinova V,
Brash HM, Flenley DC. Transient hypoxemia during sleep in chronic obstructive
pulmonary disease is not a sleep apnea syndrome. Am Rev Respir Dis 1983;128:24-29.
- Fleetham J, West P, Mezon B, Conway W, Roth T, Kryger M. Sleep, arousals,
and oxygen desaturation in chronic obstructive pulmonary disease. The
effect of oxygen therapy. Am Rev Respir Dis 1982;126:429-433.
- Fletcher EC, Miller J, Divine GW, Fletcher JG, Miller T. Nocturnal
oxyhemoglobin desaturation in COPD patients with arterial oxygen tensions
above 60 mm Hg. Chest 1987;92:604-608.
- Levi-Valensi P, Weitzenblum E, Rida Z, Aubry P, Braghiroli A, Donner
C, Aprill M, Zielinski J, Wurtemberger G. Sleep-related oxygen desaturation
and daytime pulmonary haemodynamics in COPD patients. Eur Respir J 1992;5:301-307.
- Chaouat A, Weitzenblum E, Kessler R, Charpentier C, Ehrhart M, Levi-Valensi
P, Zielinski J, Delaunois L, Cornudella R, Moutinho dos Santos J. Sleep-related
O2 desaturation and daytime pulmonary haemodynamics in COPD patients
with mild hypoxaemia. Eur Respir J 1997;10:1730-1735.
- Robin ED, Whaley RD, Crump CH, Travis DM. Alveolar gas tensions,
pulmonary ventilation and blood pH during physiologic sleep in normal
subjects. J Clin Invest 1958;37:981-989.
- Sanders MH, Newman AB, Haggerty CL, Redline S, Lebowitz M, Samet
J, O'Connor GT, Punjabi NM, Shahar E. Sleep and sleep-disordered breathing
in adults with predominantly mild obstructive airway disease. Am J Respir
Crit Care Med 2003;167:7-14.
- Catterall JR, Calverley PM, MacNee W, Warren PM, Shapiro CM, Douglas
NJ, Flenley DC. Mechanism of transient nocturnal hypoxemia in hypoxic
chronic bronchitis and emphysema. J Appl Physiol 1985;59:1698-1703.
- Hudgel DW, Martin RJ, Capehart M, Johnson B, Hill P. Contribution
of hypoventilation to sleep oxygen desaturation in chronic obstructive
pulmonary disease. J Appl Physiol 1983;55:669-677.
- Flenley DC. Clinical hypoxia: Causes, consequences, and correction.
Lancet 1978;1:542-546.
- George CF, West P, Kryger MH. Oxygenation and breathing pattern during
phasic and tonic REM in patients with chronic obstructive pulmonary
disease. Sleep 1987;10:234-243.
- Weitzenblum E, Muzet A, Ehrhart M, Ehrhart J, Sautegeau A, Weber
L. Variations nocturnes des gaz du sang et de la pression artérielle
pulmonaire chez les bronchitiques chroniques insuffisants respiratoires.
Nouv Presse Med 1982;11:1119-1122.
- Fletcher EC, Gray BA, Levin DC. Nonapneic mechanisms of arterial
oxygen desaturation during rapid-eye-movement sleep. J Appl Physiol
1983;54:632-639.
- Hudgel DW, Devadatta P. Decrease in functional residual capacity
during sleep in normal humans. J Appl Physiol 1984;57:1319-1322.
- Boysen PG, Block AJ, Wynne JW, Hunt LA, Flick MR. Nocturnal pulmonary
hypertension in patients with chronic obstructive pulmonary disease.
Chest 1979;76:536-542.
- Coccagna G, Lugaresi E. Arterial blood gases and pulmonary and systemic
arterial pressure during sleep in chronic obstructive pulmonary disease.
Sleep 1978;1:117-124.
- Fletcher EC, Levin DC. Cardiopulmonary hemodynamics during sleep
in subjects with chronic obstructive pulmonary disease. The effect of
short- and long-term oxygen. Chest 1984;85:6-14.
- Block AJ, Boysen PG, Wynne JW. The origins of cor pulmonale; a hypothesis.
Chest 1979;75:109-110.
- Tirlapur VG, Mir MA. Nocturnal hypoxemia and associated electrocardiographic
changes in patients with chronic obstructive airways disease. N Engl
J Med 1982;306:125-130.
- Flick MR, Block AJ. Nocturnal vs diurnal cardiac arrhythmias in patients
with chronic obstructive pulmonary disease. Chest 1979;75:8-11.
- Shepard JW, Jr., Garrison MW, Grither DA, Dolan GF. Relationship
of ventricular ectopy to oxyhemoglobin desaturation in patients with
obstructive sleep apnea. Chest 1985;88:335-340.
- Shepard JW, Jr., Schweitzer PK, Keller CA, Chun DS, Dolan GF. Myocardial
stress. Exercise versus sleep in patients with COPD. Chest 1984;86:366-374.
- Chaouat A, Weitzenblum E, Krieger J, Ifoundza T, Oswald M, Kessler
R. Association of chronic obstructive pulmonary disease and sleep apnea
syndrome. Am J Respir Crit Care Med 1995;151:82-86.
- National Heart Lung and Blood Institute, NIH, and World Health Organization.
Global Initiative for Chronic Obstructive Lung Disease. Global strategy
for the diagnosis, management and prevention of chronic obstructive
pulmonary disease. Executive summary. Updated 2003.
- Calverley PM, Brezinova V, Douglas NJ, Catterall JR, Flenley DC.
The effect of oxygenation on sleep quality in chronic bronchitis and
emphysema. Am Rev Respir Dis 1982;126:206-210.
- Goldstein RS, Ramcharan V, Bowes G, McNicholas WT, Bradley D, Phillipson
EA. Effect of supplemental nocturnal oxygen on gas exchange in patients
with severe obstructive lung disease. N Engl J Med 1984;310:425-429.
- Chaouat A, Weitzenblum E, Kessler R, Charpentier C, Enrhart M, Schott
R, Levi-Valensi P, Zielinski J, Delaunois L, Cornudella Ret al. A randomized
trial of nocturnal oxygen therapy in chronic obstructive pulmonary disease
patients. Eur Respir J 1999;14:1002-1008.
Last updated: 8
June 2005Created |
 |
| Legal
notice |
© GE
Healthcare 2008
ISSN 1795-6269 (Web)
ISSN 1795-6277 (CD) |
Webmaster |
|
 |
|